| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |

From the Associated Regional and University Pathologists Institute for Clinical and Experimental Pathology
*
and the Department of Pathology,
University of Utah School of Medicine, Salt Lake City, Utah
| Abstract |
|---|
|
|
|---|
| Introduction |
|---|
|
|
|---|
Fluorescence monitoring of PCR product melting profiles permits the detection of DNA mutations in solution, without the need for gel or column separation.8 Fluorescently labeled probe-based methods such dual hybridization,9 exonuclease (TaqMan),10 or hairpin (Molecular Beacon)11 probes may be used for mutation detection, but only for the bases covered by the probe; thus, they are not amenable to mutation scanning. Mutational scanning requires methods that can detect aberrations over larger regions. A number of methods have been described which use SYBR Green I and melting curve analysis.6, 12, 13 However, these methods required either primer modification with GC-rich sequences and/or extensive post-amplification manipulation before melting analysis. Additional homogeneous methods for scanning by melting analysis include the use of labeled primers14 or multiple overlapping probes to cover the length of the amplicon.15 Recently, a mutation scanning method using high-resolution melting analysis (HRMA) has been described.16 HRMA can detect mutations over the length of an amplicon without the need for gel or column separation. This closed-tube method employs the dsDNA binding dye LCGreen I and obviates the need for fluorescently labeled primers or probes. To date, the utility of this method has only been investigated for the detection of point mutations, single nucleotide polymorphisms, or similarly subtle variations in nucleic acid sequences.
In this study, we examine the feasibility of using HRMA for the detection of large sequence aberrations, using internal tandem duplications (ITD) in the juxtamembrane domain of the FLT3 gene as a model system. ITD mutations in FLT3 have been identified in approximately 20% of acute myelogenous leukemias (AML).17 These mutations occur in exon 14 and intron 14 of the FLT3 gene but the exact location and length of the mutations varies between patients. We screened DNA from 62 AML samples by HRMA for the presence of ITD mutations. Samples were also analyzed by capillary electrophoresis (CE)-based fragment analysis and TGCE. All ITD mutations were further confirmed by sequencing. Our studies show a 100% correlation between HRMA and the other methods, thereby illustrating the utility of HRMA for the detection of large sequence aberrations.
| Materials and Methods |
|---|
|
|
|---|
Polymerase Chain Reaction
DNA samples were amplified for the FLT3 locus using the previously described primers 11F, 5'-GCAATTTAGGTATGAAAGCCAGC-3', and 12R, 5'-CTTCAGCATTTTGACGGCAACC-3'.18
Analysis by high-resolution melting and TGCE required product that was free of polymerase-dependent errors, so high fidelity amplification was used. One hundred nanograms of DNA was amplified in a 50 µl reaction containing 0.3 µmol/L each of each primer and 1X High Fidelity Master Mix (Roche Molecular Biochemicals) (containing Taq and Tgo DNA polymerases, 0.2 mmol/L each dNTP, and 1.5 mmol/L MgCl2). The amplification was performed on the GeneAmp PCR System 2700 (Applied Biosystems, Foster City, CA) and entailed an initial denaturation of 94°C for 2 minutes, followed by 30 cycles of denaturation at 94°C for 30 seconds, annealing at 64°C for 30 seconds, and extension at 72°C for 1 minute, with a final extension at 72°C for 5 minutes. DNA samples for FLT3 mutation analysis by capillary electrophoresis were amplified using a modified version of the above protocol with the 11F primer 5'-labeled with 6-FAM.
Amplicon Detection
High-Resolution Melting Analysis
All FLT3 amplicons were subjected to HRMA. LCGreen I dye (Idaho Technology, Salt Lake City, UT) was added to each amplicon at a final concentration of 10 µmol/L. To facilitate heteroduplex formation, 10 µl of each sample was heated in a glass capillary on the LightCycler (Roche Molecular Biochemicals) at 95°C for 1 minute and then cooled to 40°C for 2 minutes with a temperature transition of 20°C/second. Samples were then heated at 0.2°C/second from 60 to 90°C on a prototype of the Idaho Technology HR-1 instrument. This glass capillary-based single-sample high-resolution melting instrument collects fluorescence data at a rate of approximately 50 data points per 1°C.
Results were analyzed as fluorescence (F) versus temperature (T) graphs. The data were also analyzed using normalization, temperature shifting, difference plots, and derivative melting curves, as previously described.16 Briefly, for normalization, temperature ranges on each side of the melting transition were chosen and the data points for a given sample were scaled between 0 and 100% fluorescence. For temperature shifting, the melting curves of the samples were superimposed over the same temperature range. This aided in the differentiation between wild-type and mutant samples by emphasizing curve shape in conjunction with melting temperature. Difference plots were generated by choosing a wild-type control as the baseline and subtracting the fluorescence of each sample relative to this baseline. Derivative plots were visualized as the negative derivative of the fluorescence relative to the temperature (-dF/dT) versus temperature.
Capillary Electrophoresis-Based Fragment Analysis
FLT3 PCR products were run on the ABI PRISM 3100 Genetic Analyzer (Applied Biosystems) using a 50-cm array with Performance Optimized Polymer-6 (POP-6). One µl of PCR product was mixed with 8.5 µl formamide and 0.5 µl ROX500 internal size standard. One µl of this mixture was electrokinetically injected at 10 volts for 1 second. Electrophoresis was performed at 1500 volts at 55°C with a data delay time of 1120 seconds and a run time of 4000 seconds. Data were analyzed with the ABI Filter Set D using the 3100 Data Collection software, version 1.1 (Applied Biosystems).
Temperature Gradient Capillary Electrophoresis
Temperature gradient capillary electrophoresis was performed post-amplification for 36 of the AML patient samples and for the MV411 cell line. Following amplification, samples were heated to 94°C for 5 minutes and then cooled to 4°C for a minimum of 5 minutes for heteroduplex formation. Samples were injected at 3 kV for 20 seconds and assayed on the SpectruMedix Reveal System (State College, PA) with a temperature range of 45 to 60°C and a ramp time of 23 minutes. Results were analyzed using the SpectruMedix Revelation software. Samples were genotyped by comparison of the electropherograms to those of the wild-type control. Samples were scored as positive for an ITD mutation if they exhibited peaks in addition to the pattern of the wild-type control.
Analytical Sensitivity
DNA from the MV411 cell line was serially diluted into placental DNA. Dilutions of 1X, 1:2, 1:5, 1:10, and 1:20 were amplified using the FLT3 primer set and assessed using both high-resolution melting and capillary electrophoresis.
Sequencing
PCR amplicons from samples with ITD mutations were sequenced to confirm the presence of and to characterize the internal duplication. The amplicons were separated by electrophoresis on 2% or 4% agarose gels and the ITD bands were purified using the QIAquick Gel Extraction Kit (QIAGEN Inc., Valencia, CA). Automated DNA sequencing of PCR products was performed using dideoxynucleotide termination chemistry and the ABI PRISM 3100 Genetic Analyzer (Applied Biosystems).
| Results |
|---|
|
|
|---|
|
Results were also visualized as difference plots and derivative plots (-dF/dT versus T graphs) of the melting curve data (Figure 1 D and E)
. Using the fluorescence difference plots, the curves for the wild-type samples were clustered around the baseline while the curves for the mutant samples were always seen in the positive scale regardless of which wild-type sample was used as the control for the baseline. The derivative plots of wild-type samples showed a single peak. Samples with ITD mutations exhibited either a single peak with a shoulder or two peaks.
The MV411 cell line and five of the samples containing ITD mutations exhibited multiple melting domains. The five AML samples all had significant populations of the ITD allele relative to the wild-type allele as determined by the peak heights in the CE electropherograms. All other ITD samples resulted in broad melting transitions.
Capillary Electrophoresis-Based Fragment Analysis
Wild-type samples yielded a single peak at 329 bases and ITD mutant samples yielded one or two peaks greater than 329 bases (see Figure 2
). The wild-type peak was present in all clinical samples. However, the relative intensity of the ITD allele to the wild-type allele varied greatly. The MV411 cell line exhibited a single peak at 359 bases.
|
|
Analytical Sensitivity
The ITD mutation was detectable at a dilution of 1:5 using HRMA and at a dilution of 1:10 using CE-based fragment analysis.
Sequencing Analysis
ITD mutations were confirmed by DNA sequencing of PCR products. Internal tandem duplication PCR products ranged in size from 335 to 431 bases in length, representing duplications of 6 to 102 bases. One sample contained two ITD products with lengths of 389 and 431 bases. Although the length and exact location of the insertions varied between samples, all ITD mutations resulted in in-frame shifts. An example of the sequence associated with an internal tandem duplication is illustrated in Figure 4
.
|
| Discussion |
|---|
|
|
|---|
The employment of solution-based or real-time fluorescence methods for the detection of structural variations in DNA sequences has entailed the utilization of sequence-specific fluorescently labeled probes or non-specific DNA binding dyes.19, 20, 21, 22 The sequence-specific DNA probe-based methods are generally limited by their inability to detect sequence alterations that occur outside the coverage area of the "reporter" oligonucleotide probe. By comparison, the non-specific dsDNA binding dye-based techniques for mutational scanning require somewhat extensive post-amplification manipulation for the reliable detection of variations between two DNA sequences. In this regard, we previously demonstrated the feasibility of detection of single-base substitutions in amplicons > 200 bp in length, by incorporating GC-clamps into one of the PCR primers used for the amplification of the DNA segment of interest, and performing a post-amplification melting protocol which involved the incorporation of urea into the reaction mixture.6 This procedure is labor intensive, and subject to the achievement of variable results due to the multiple post-amplification steps required for scanning of DNA for sequence alterations. Discrimination of mutated sequences has also been achieved using 5'-fluorescently labeled primers in PCR.14 With this approach, genotyping of amplicons of up to 300 bp in size was successfully performed, but was limited by the requirement that the sequence alteration interrogated be within a melting domain that included the fluorescently labeled primer.14 Hence, the development of simple closed-tube fluorescence-based methods that discriminate between related sequences would be a desirable improvement in molecular diagnostics.
Post-amplification analysis of amplicons by gradual melting of the DNA sequences in combination with acquisition of multiple fluorescence data points (50 per second) permits detailed evaluation of the status of homo- and heteroduplexes, and strand dissociation states during dsDNA strand denaturation.16 The HRMA instrument collects more fluorescence data per degree Celsius than other instrumentation such as the LightCycler or ABI PRISM 7900HT. This approach and the utilization of a dsDNA binding dye such as LCGreen I, which does not significantly alter the relationship between DNA duplexes in solution, constitute the underlying principles inherent to the feasibility of high-resolution melting in nucleic acid genotyping. Notwithstanding the promising results demonstrated in the genotypic discrimination of sequences with germ-line alterations comprising single-base substitutions, the utility of identification of larger sequence aberrations and somatically acquired mutations has not been previously evaluated.
In this study, we have demonstrated the utility of high-resolution melting in conjunction with the generic dsDNA binding dye LCGreen I in the identification of sequence alterations larger than point mutations. While the majority of our experiments incorporated LCGreen I post-amplification, we have demonstrated that this dye is compatible with PCR and its inclusion at the beginning of PCR does not adversely affect the ability to perform HRMA. In addition to being PCR-compatible, the use of LCGreen I is advantageous because unlike SYBR Green I, LCGreen I does not appear to undergo significant repartitioning within the double-stranded DNA strand during post-amplification melting.16 This permits finer resolution of melting profiles and thus improves the ability to detect nucleic acid sequence variations. Using internal tandem duplications of the FLT3 gene as a model target, we have demonstrated that high-resolution melting using LCGreen I can reliably detect DNA insertions ranging from 6 to 102 bases within amplicons ranging from 335 to 431 bp in length. Each ITD was detected as a reproducible variation in the shape of the F versus T curves of the samples containing the ITDs when compared to the curves of the samples with the wild-type sequence. The difference in the curve-shape is attributable to heteroduplex formation between the mutated sequence contributed by the abnormal alleles in the tumor DNA and wild-type sequence contributed by a normal allele in the tumor DNA, or from DNA from contaminating normal cells.
Interestingly, we also reproducibly detected the presence of a homozygous FLT3 ITD in an acute myelogenous leukemia-derived tumor cell line (MV411) within which no wild-type sequence was demonstrable by CE, TGCE, or DNA sequencing. The ITD in this sample was composed of a 30-bp sequence insertion, and yielded an interesting F versus T profile with two distinct melting transitions. Analysis of the sequence within the ITD revealed a likelihood for the formation of misannealed heteroduplexes in configurations following re-annealing of the DNA strands in the homozygous 30-bp ITD present in the MV411 DNA.
The detection of FLT3 ITD mutations by HRMA was 100% concordant with the results obtained by CE, TGCE, and sequencing for the same samples. HRMA was advantageous because samples could be screened within 2 minutes post-amplification. HRMA was effective at detecting mutations across the length of the amplicon since the position of the mutation varied between samples. Theoretically, sequence aberrations such as deletions and random insertions would be similarly detectable in heterogeneous samples with wild-type and mutated species. Although only sequencing can provide exact characterization of each mutation, HRMA was able to both rapidly and accurately determine which samples contained mutations. We found the primary drawback of the HRMA method to be its inability to detect small quantities of mutant DNA species in an overwhelming background of wild-type DNA species. This is particularly important in the identification of somatically acquired mutations, wherein the tumor cells account for only a small proportion of the total cell population. With the dilutional series study, we found that ITD alleles were only detected in samples when comprising up to 20% of the entire target DNA content analyzed. Capillary electrophoresis could distinguish ITD populations present at 10%. Both systems are limited by the preferential amplification of the dominant species.
In conclusion, we successfully identified somatic mutations of the FLT3 gene in 13 of 62 (21%) AML patient samples assessed using HRMA. These results correlated 100% with CE, TGCE, and sequencing results. Our studies show that HRMA has the advantage of screening each sample for mutations within 2 minutes post-amplification. We found the main limitation of HRMA to be its inability to identify mutations when the relative abundance of mutated cells was less than 20% of the total cell population. Future developments to increase the frequency of fluorescence acquisition may increase sensitivity and allow the system to discriminate sequence variations in longer DNA segments. Nevertheless, our studies illustrate that HRMA in its current design can be used to quickly and accurately screen DNA samples for large sequence aberrations, including FLT3 ITDs.
| Acknowledgments |
|---|
| Footnotes |
|---|
Supported by the ARUP Institute for Clinical and Experimental Pathology, Salt Lake City, UT.
Accepted for publication April 16, 2004.
| References |
|---|
|
|
|---|
-globin gene loci in normal and Hb Lepore DNA: direct determination of gene linkage and intergene distance. Cell 1978, 15:25-41[Medline]
This article has been cited by other articles:
![]() |
A. J. Stephens, J. Inman-Bamber, P. M. Giffard, and F. Huygens High-Resolution Melting Analysis of the spa Repeat Region of Staphylococcus aureus Clin. Chem., February 1, 2008; 54(2): 432 - 436. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. L. Kennerson, T. Warburton, E. Nelis, M. Brewer, P. Polly, P. De Jonghe, V. Timmerman, and G. A. Nicholson Mutation Scanning the GJB1 Gene with High-Resolution Melting Analysis: Implications for Mutation Scanning of Genes for Charcot-Marie-Tooth Disease Clin. Chem., February 1, 2007; 53(2): 349 - 352. [Abstract] [Full Text] [PDF] |
||||
![]() |
J.-C. Cheng, C.-L. Huang, C.-C. Lin, C.-C. Chen, Y.-C. Chang, S.-S. Chang, and C.-P. Tseng Rapid Detection and Identification of Clinically Important Bacteria by High-Resolution Melting Analysis after Broad-Range Ribosomal RNA Real-Time PCR Clin. Chem., November 1, 2006; 52(11): 1997 - 2004. [Abstract] [Full Text] [PDF] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |