| Abstract |
|---|
|
|
|---|
| Introduction |
|---|
|
|
|---|
There are a variety of molecular methods available for CPN detection. These methods, including nested PCR and touchdown enzyme time release (TETR) PCR, have improved sensitivity over traditional methods of CPN detection.5, 7, 8, 9 However, these methods also have their limitations. Traditional PCRs typically rely on post-PCR processing for the determination of a result. In addition, recent reports suggest that there are inherent problems in using traditional PCRs for detection of CPN due to lack of standardization between laboratories that can result in increased variation.5, 10, 11, 12
Real-time PCR offers the combination of amplification and detection in one step, eliminating the need for post-PCR processing. Real-time PCR also offers improved sensitivity over traditional PCR as well as the ability to quantify the amount of amplicon generated from a reaction.13 Quantification may prove useful in differentiating previous, acute, and chronic CPN infections.
In this study, a previously published primer set CPN90 and CPN91 targeting a 197-bp region of the 16S rRNA gene of CPN was used with a Taqman probe (CPNTM) and the Roche Lightcylcer platform to develop a method for real-time detection of CPN from peripheral blood mononuclear cells (PBMCs) and respiratory specimens. This method has several advantages over traditional PCRs including decreased labor by eliminating the need for post-PCR processing of the sample and comparable, if not increased, sensitivity and specificity.
| Materials and Methods |
|---|
|
|
|---|
Specimen Preparation
Three different sets of specimens were used for assay validation. The first set consisted of two blinded panels of pre-extracted CPN DNA from culture (N = 70). The second set consisted of archived specimens consisting of a CPN dilution series, CPN-spiked porcine aortal tissue and endarterectomy specimens (N = 87),12
while the final set consisted of prospectively collected PBMCs from a heart disease study (N = 198). The first set of samples from culture, which were previously DNA extracted, was thawed and subjected to the CPNTM PCR assay. The second and third set were extracted using the Roche MagNA Pure LC robotic instrument in an identical manner to the standards, and subjected to the CPNTM PCR assay. Due to the variation in sample type and age, comparison of results between the sample sets was inappropriate, therefore, individual sensitivities and specificities for each set were calculated.
Primer and Probe Design
A previously published set of primers CPN90 (5'ggtctcaaccccatccgtgtcgg 3') and CPN91 (5' tgcggaaagctgtatttctacagtt 3') was used for the CPNTM PCR assay.7
A probe (CPNTM, 5'6FAM-atgccgcctgaggagtacactcgcaa-tamra 3'), using Taqman technology, was developed for detection of a 197-bp region of the 16S rRNA gene amplicon (TIB Molbiol, Adelphia, NJ).
PCR Conditions
PCR reactions contained 18 µl of master mix and 2 µl of template DNA. The final reaction mixture contained: 4 mmol/L MgCl2, 20 pM of the primers CPN90 and CPN91, 20 pM of the CPNTM probe, and 1X LC FastStart DNA Master Hybridization Probes buffer based on the manufacturers recommendations (Roche). Cycling conditions included a pre-incubation step of 4 minutes at 95°C followed by 45 cycles of 95°C for 0 seconds, 65°C for 10 seconds, and 72°C for 5 seconds. The amplification protocol was followed by a cooling period of 40°C for 30 seconds. Data collections were performed during extension and were monitored through the F1 channel of the instrument. Data analyses were performed using the Fit Points method through the Lightcycler software to minimize noise.
Nested PCR and TETR PCR were performed as previously described, and gel electrophoresis was used to determine the positivity of the samples.7, 8 TETR PCR targeted the 16S rRNA gene that is the same target as the CPNTM PCR assay, while nested PCR targeted the ompA gene, thus providing an alternative target for assay comparison.
Reproducibility Study
Standards (strain A03) of 400 IFU/PCR reaction, 40 IFU/PCR reaction, 4 IFU/PCR reaction and 0.4 IFU/PCR reaction were used to determine the reproducibility of the CPNTM PCR assay. Standards were serially diluted, extracted as described above, and subjected to the CPNTM PCR assay by two different technicians on two separate days. Standards were tested with eight replicates per standard. Variability is shown as the SD (SD). Statistical and regression analyses were carried out with STATA Version 7.0 (Stata Corp., College Station, TX) and Sigma Plot Software (SPSS Inc., Chicago, IL).
Specificity Testing
Both Chlamydia psittaci (VR125) and Chlamydia trachomatis (Serovar E) were tested to determine any potential cross-reactivity between the primers and probes with these organisms. Both Chlamydia psittaci and Chlamydia trachomatis samples had concentrations of 40 IFU/PCR reaction.
Sample Testing
A processing negative control, positive controls, consisting of a group of CPN standards (strains A03 or AR39 or AR388) of 400 IFU/PCR reaction, 40 IFU/PCR reaction, 4 IFU/PCR reaction, and 0.4 IFU/PCR reaction were included in all runs of tested samples. True positives were defined as positive either by a known result provided by an investigator based on culture results in the case of the standardized panels, or positive by two of the three PCRs used, ie, CPNTM PCR, nested PCR, or TETR PCR. Discrepant analysis was not performed in most cases due to limited sample volume.
| Results |
|---|
|
|
|---|
Reproducibility Studies
On two separate days, two different technicians tested a panel of four standards with concentrations of 400 IFU per PCR reaction, 40 IFU/PCR reaction, 4 IFU/PCR reaction, and 0.4 IFU/PCR reaction. Results, including calculated concentrations and cycling times/crossing points (Ct) are indicated in Table 1
, as well as in Figure 1
. To assess the reproducibility, multiple linear regression was performed to create standard curves corresponding to the two runs. Ct was regressed on the log concentrations generated by Technician 1 and Technician 2 with a model allowing the intercepts and slopes of the standard curves corresponding to the two separate runs to differ. Analysis showed the fitted lines to be statistically different (F (2.53) = 5.25, P = 0.0091), although the overall variation in end point detection was not different for practical purposes (Figure 1)
.
|
|
Sample Testing
A total of 355 samples were tested over 23 runs in which positive control standards of 400 IFU/PCR reaction, 40 IFU/PCR reaction, 4 IFU/PCR reaction and 0.4 IFU/PCR reaction (strains A03, AR39, or AR388) and negative controls were used in every run to determine the Ct for positive samples, and account for background fluorescence.
The sensitivity and specificity of the CPNTM PCR assay were determined by comparing results to nested PCR, TETR PCR, or the known result for the sample from previous study.12
Overall, the assay had a sensitivity of 88.5% (69 of 78) and a specificity of 99.3% (275 of 277). Tested samples included two panels of blinded samples from culture (N = 70), archived specimens consisting of a CPN dilution series, CPN spiked porcine aortal tissue and endarterectomy specimens (N = 87), and PBMCs from clinical patients (N = 198). Sensitivities and specificities were also calculated separately for each group due to variations in sample type, age of sample, and extraction method. It should also be noted that because there was variation in the treatment of the different sample sets, the results are not necessarily comparable across the sets. The sensitivity and specificity for the standardized panels were 94.4% (51 of 54) and 100% (16 of 16), respectively, for the archived specimens 75% (18 of 24) and 100% (63 of 63) respectively, and for the PBMCs zero positives were detected, and the specificity was 98.9% (196 of 198) (Tables 2
and 3)
.
|
|
| Discussion |
|---|
|
|
|---|
The CPNTM PCR assay combines the sensitivity and specificity of the CPN90 and CPN91 primer set with the speed, accuracy, and ease of real-time PCR technology. This assay is rapid, taking less than 3 hours from DNA extraction to completion of the PCR and sensitive, with a detection limit of 4 to < 1 IFU/PCR reaction.
The overall sensitivity and specificity of the CPNTM PCR assay is comparable to other real-time methods available, including ompA-based PCRs.5 Additionally, this method is comparable to nested PCR and TETR PCR, the gold standards used in development of this assay.7, 8 Based on other inter-method comparisons, the CPNTM PCR assay is comparable to recently available commercial research kits for CPN detection (PCR research assay, Abbott Laboratories).5
The sensitivity of the CPNTM PCR assay varied depending on the sample set. Because the samples differed in sample type, age, and extraction method, it was reasonable to calculate separate sensitivities and specificities, as well as the overall. The variation in treatment between the different sample sets makes comparison of the differing results difficult. However, the CPNTM PCR assay was still comparable to other available methods for CPN detection in terms of the overall sensitivity and specificity. For sample set 1 that consisted of two blinded panels of samples from culture, the sensitivity (94.4%) and specificity (100%) of the CPNTM PCR assay was comparable to other real-time methods available.5 However, for sample set 2 that consisted of archived specimens consisting of a CPN dilution series, CPN-spiked porcine aortal tissue and endarterectomy specimens, the sensitivity of the assay was much lower (75%) than the overall sensitivity (88.5%), and much lower than other real-time methods for CPN detection.5 The lower sensitivity is most likely due to DNA degradation through multiple freeze-thaw processes and time. For sample set 3 that consisted of prospectively collected PBMCs, the sensitivity of the CPNTM PCR assay was non-calculable because there were no positives detected by the gold standard assays. The two positives that were identified by the CPNTM PCR assay were not confirmed by nested PCR or by TETR PCR. Although these two positives were not confirmed, it is possible that the CPNTM PCR is more sensitive, and that the two unconfirmed positives were true positives. Alternatively, the two unconfirmed positives could represent false positives, or potentially contaminated samples.
Cost is a limitation of the assay, as the equipment used in the assay can cost up to $50,000, and PCR reactions can cost approximately $8.00 per reaction without accounting for technician time. In addition, the cost of automated nucleic acid extraction can be considerable, as the instrument alone costs upwards of $80,000. However, despite this limitation, this real-time method appears to be a practical method of accurate, high throughput detection of CPN, and should serve to improve the reliability of diagnostic testing for CPN while minimizing manual processing, sample handling, and post-amplification manipulation. Future studies with this method should include prospective analysis of a large sample group, treated and extracted the in the same manner with comparison to a truly accurate gold standard.
| Acknowledgments |
|---|
| Footnotes |
|---|
Accepted for publication February 10, 2004.
| References |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
A. Kutlin, P. M. Roblin, S. Kumar, S. Kohlhoff, T. Bodetti, P. Timms, and M. R. Hammerschlag Molecular characterization of Chlamydophila pneumoniae isolates from Western barred bandicoots J. Med. Microbiol., March 1, 2007; 56(3): 407 - 417. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Morozumi, E. Nakayama, S. Iwata, Y. Aoki, K. Hasegawa, R. Kobayashi, N. Chiba, T. Tajima, K. Ubukata, and the Acute Respiratory Diseases Study Group Simultaneous Detection of Pathogens in Clinical Samples from Patients with Community-Acquired Pneumonia by Real-Time PCR with Pathogen-Specific Molecular Beacon Probes J. Clin. Microbiol., April 1, 2006; 44(4): 1440 - 1446. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. J. Kohlhepp, J. Hardick, and C. Gaydos Chlamydia pneumoniae in Peripheral Blood Mononuclear Cells Isolated from Individuals Younger than 20 Years or Older than 60 Years J. Clin. Microbiol., June 1, 2005; 43(6): 3030 - 3030. [Full Text] [PDF] |
||||
![]() |
R. B. Raggam, E. Leitner, J. Berg, G. Muhlbauer, E. Marth, and H. H. Kessler Single-Run, Parallel Detection of DNA from Three Pneumonia-Producing Bacteria by Real-Time Polymerase Chain Reaction J. Mol. Diagn., February 1, 2005; 7(1): 133 - 138. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. M. Ieven, V. Y. Hoymans, J. M. Bosmans, D. Sander, K. Winbeck, J. Klingelhofer, T. Etgen, and B. Conrad Early Carotid Atherosclerosis and Chlamydia pneumoniae Seropositivity: Are There Arguments to Treat With Antibiotics? * Response Circulation, August 17, 2004; 110(7): e74 - e75. [Full Text] [PDF] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |